Pt100, Pt1000 and thermocouples

Check a sensor with a meter: enter the measured resistance or voltage and see the temperature. It also works the other way, for example when simulating a sensor with a calibrator.

RTD: Ω → °C
Ω
Ω

Result

Temperature24.988°C
Class B tolerance
± 0.425 °C
Sensitivity
0.3879 Ω/°C
R / R₀
1.0973

R = R₀ · (1 + A·T + B·T²)

A = 3.9083·10⁻³ B = −5.775·10⁻⁷ C = −4.183·10⁻¹²

Pt100 table
T [°C]R [Ω]class A ±°Cclass B ±°C
−20018.52010.551.3
−10060.25580.350.8
−5080.30630.250.55
01000.150.3
25109.7350.20.425
50119.3970.250.55
100138.5060.350.8
150157.3250.451.05
200175.8560.551.3
300212.0520.751.8
400247.0920.952.3
500280.9781.152.8
600313.7081.353.3
850390.4811.854.55
Embed on your site

Paste this code into your page’s HTML. The tool sets its own height, and the link to Kalkio has to stay visible.

Calculator by Kalkio · kalkio.app

Pt100 and Pt1000 resistance thermometers

A Pt100 is 100 Ω at 0 °C and a Pt1000 is 1000 Ω. Resistance versus temperature follows the Callendar–Van Dusen equation from IEC 60751. A Pt100 changes by about 0.385 Ω per degree, so 1 Ω of lead resistance in a 2-wire circuit gives an error of roughly 2.6 °C. That is why industrial installations use 3- and 4-wire connections.

Thermocouples

A thermocouple produces a voltage that depends on the temperature difference between the measuring junction and the cold junction. ITS-90 tables and formulas assume the cold junction is at 0 °C. When the terminals are at another temperature, add E(T₀), the thermocouple’s voltage at the terminal temperature, to the measured voltage. That is what cold-junction compensation in a transmitter does.

Thermocouple voltages are calculated from the NIST ITS-90 reference functions (NIST SRD 60).

Common questions

What is the resistance of a Pt100 at 20 °C?

107.79 Ω. At 100 °C it is 138.51 Ω.

How do I check a thermocouple with a multimeter?

Measure the mV at the terminals, measure the terminal temperature and enter both values in “mV → °C” mode.

Why does type B start at 50 °C?

Below about 50 °C a type B thermocouple’s voltage is almost zero and ambiguous, so temperature can’t be read from it.